5.5. Gravity Compensation Function
□ If the scale is used at the caiibraiion location, it is not necessary to perform this function.
it may cause a weighing error, if there is a difference of gravity acceleration
between the installed location and calibration location. This function specifies
these gravity accelerations and corrects span error.
Caution The decimal point is not displayed in the function. Ex, 313B = 9.798 m/s^
Step 1 Press the ICALl key in normal weighing mode. The Normai weighing mode
I C3L in is displayed for two seconds. Press the 1M+
key to enter the gravity compensation function.
If you want to cancel the current procedure, press
and hold the |NEi/GR0SS| key and press the |CAL| key. Then,
all parameters are not changed and the calibration
mode is finished. Press the lOperate/Standb^ key to turn the
display off after displaying CRnCEL
Step 2 The parameter
01 xxxx is displayed. Enter the
>
gravity acceleration of the calibration location. The
parameter xxxx is the gravity acceleration.
key
..........
The key to select a figure.
key
...........
The key to select a number.
key
..........
The key to store the new gravity
acceleration and proceed to step 3.
A
mode! key
.........
The key not to change the value and
proceed to step 3.
Step 3 The parameter OB xxxx is displayed. Enter the
gravity acceleration of the installed location. The
parameter xxxx Is the gravity acceleration.
>n key
............
The key to select a figure.
7T| key
............
The key to select a number.
key...
........
The key to store the new gravity
acceleration and proceed to step 4.
mode! key
.........
The key not to change the value and
proceed to step 4.
Step 4 Now [ ¿W xxxx I is displayed. Press the ICAl] key to
store the parameters. The I End I is displayed.
Proceed to steps.
Step 5 Press the loperate/Siandby key to turn display off.
AD-4329 Weighing Indicator Page 23
2 seconds
Calibrated location
Installed location
CAL
[Ri in
d”0P
M4*
6 I 3000
o
ZERO
>
-V|V
TARE
A
0 t 3138
Q
PRINT
08 3000
ZERO
>
TARE
A
08 3800
Q
PRINT
0
1 3198
CAL
End
I/O
0PR/ST8
Turn display off
5.5. Gravity CompensaBon Function