4.1.2. Rear Panel Description
No.
Name Description
15 Power connector
Use accessory power cord.
Please Confirm that the iocal voltage and receptacle type are
correct for your scale.
16 Grounding terminal
17 RS-232C interface
Refer to "12. RS-232C Interface" for the detail.
18 Blank panel
Option space (for relay output).
19 Fuse holder
Use 0.5A time lag fuse for AC1OOV to AC120V.
Use 0.316A time lag fuse for AC200V to AC240V.
20 External input
Seven control inputs selected function by the F functions.
Refer to FI 0 to 16.
21 Serial output
Printer port. Refer to "11. Standard Serial Output" for the
details.
22
Load cell connector
Connect accessory load cell plug.
4.1.3. Other Displays and Symbols
standby display.
Zero error when turning display on. If the I mode I key is
pressed, a current weighing value may be displayed.
Blank
Decimai point
Over load display. Remove any load from the load cel!
immediately. It may cause damage to the indicator.
r r
lг
Example of an error display.
AD-4329 Weighing Indicator
Page 9
4. Description of Panels and symbols