Ad Weighing Indicator AD-4329 Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
Calibrate the scale, if It is moved to another location or the environment
'has changed.
It Is not necessary to set the gravity acceleration correction, when calibrating
the scale with a calibration mass at the place where the scale is used.
Enter the stable weighing data while the MD mark is turned off. If
unstable data Is used, it may cause a weighing error. Arrange the
condition using the FOO filter function.
The span calibration needs the zero calibration data. We recommend that
you perform the span calibration Immediately after the zero calibration.
Perform the digital linearization function immediately after the zero
calibration. And perform the span calibration continuously.
Select a decimal point and weighing unit at the CF function CFOO, CF 01.
If you use, the dual or triple range function of the multi-interval scale,
perform the “Range Function", “Zero Calibration" and "Span Calibration".
5.2. Calibration Procedure
5.2.1. The Whole Procedure
Essential Items
Step A Preparation for calibration.
Step B Minimum division.
Step C Maximum capacity.
Step B Zero calibration.
Step G Span calibration
Optional Items
Step D Weighing range function,
i Perform this function just
before zero calibration and
span calibration.
Step F Digital linearization function.
Perform this function after the
zero calibration immediately.
And perform the span
calibration continuously.
Step H Gravity compensation function.
Perform this function after span calibration.
5. Calibration Page 12
AD-4329 Weighing Indicator
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 55 56

Comentarios a estos manuales

Sin comentarios