5.3. Weighing Range Function
The weighing range function can select "single range", "dual range” and "triple range".
Specify each weighing interval (division) for the multi-interval instrument Each
........
weighing interval is displayed according to a net value or gross value.
Caution a Using a single range, It 1^ not necessary to perform this function.
□ Perform the zero and span calibration after this function.
Example 1 The gross display.
Specified parameters:
First range
Second range
Third range
Display
0kg to 20kg:
20kg to 50kg:
60kg to 100kg:
Range = 20.00kg, division 0.01 kg
Range = 50.00kg, division 0.02kg
Range - 100.00kg (maximum capacity), division 0.1kg
The first range, division 0.01 kg.
The second range, division 0.02kg,
The third range, division 0.1 kg.
Division
Division
Division
0.1kg
0,02kg
I 1 1 It
0.01kg
TiirinTiniTi
■lili
0kg
20kg
50kg
100kg Gross
Weiging value
Example 2 The net display using a 40kg tare value.
Specified parameters:
The same parameters as example 1.
Display
-40kg to -20kg :
"20kg to 20kg :
20kg to 50kg :
50kg to 60kg :
The second range, division 0.02kg.
The first range, divisiion 0.01 kg.
The second range, division 0,02kg.
The third range, division 0.1 kg.
Division
Division
0.1kg
-40kg
Division
0.02kg
1
Division
0.01kg
iiiinrríTiTm
0.02kg
lilillitlllilill
1 I1Ì11
nrminif
-20kg 0kg
0kg
20kg
50kg 60kg Net
30kg
Tare vaiue
40kg
60kg
100kg Gross
Weighing value
5.3. Weighing Range Function